Modules and Deck Guide

Module compatibility is both an API question and a physical hardware question. Confirm the robot model, module generation, caddy or adapter, firmware, deck location, and labware before authoring commands.

Robot Compatibility

Hardware Flex OT-2 Protocol API load name
Absorbance Plate Reader Yes No absorbanceReaderV1
Flex Stacker Yes No flexStackerModuleV1
Heater-Shaker GEN1 Yes Yes heaterShakerModuleV1
Magnetic Block GEN1 Yes No magneticBlockV1
Magnetic Module GEN1/GEN2 No Yes magnetic module / magnetic module gen2
Temperature Module GEN2 Yes Yes temperature module gen2
Thermocycler GEN2 Yes Yes thermocyclerModuleV2

Notes:

See the current Flex module list and OT-2 module list before relying on this snapshot.

Deck Models

Flex

trash = protocol.load_trash_bin("A3")
chute = protocol.load_waste_chute()  # D3 only; do not load both into D3

Load only the fixtures physically installed for the run.

OT-2

Module Loading

heater_shaker = protocol.load_module(
    module_name="heaterShakerModuleV1",
    location="D1",
)

Loading a powered module makes it a run requirement. The App will prevent the run from starting unless the expected connected module is available.

Load labware through the module:

plate = heater_shaker.load_labware(
    "corning_96_wellplate_360ul_flat",
    label="Mixing Plate",
)

For a separate adapter:

temperature_module = protocol.load_module(
    "temperature module gen2",
    "D3",
)
adapter = temperature_module.load_adapter(
    "opentrons_96_well_aluminum_block"
)
plate = adapter.load_labware(
    "opentrons_96_wellplate_200ul_pcr_full_skirt"
)

The API generally cannot prove that every unusual module, adapter, and labware combination is physically valid. Check the official compatibility list.

Temperature Module

temperature_module.set_temperature(celsius=4)
# Pipetting or incubation commands...
temperature_module.deactivate()

set_temperature() blocks until the target is reached. In API 2.27+, start_set_temperature() can run concurrently and returns a task. Wait for the task before relying on the target temperature.

Operational checks:

Heater-Shaker

Typical sequence:

heater_shaker.close_labware_latch()
heater_shaker.set_and_wait_for_temperature(37)
heater_shaker.set_and_wait_for_shake_speed(1000)
protocol.delay(minutes=5)
heater_shaker.deactivate_shaker()
heater_shaker.deactivate_heater()

The exact temperature method name depends on API behavior. Current context methods include set_target_temperature(), set_and_wait_for_temperature(), wait_for_temperature(), set_and_wait_for_shake_speed(), deactivate_shaker(), and deactivate_heater().

Always:

API 2.27+ provides nonblocking shake and temperature operations for concurrent work. Keep task handles and wait before a dependent step.

Thermocycler

thermocycler = protocol.load_module("thermocyclerModuleV2")
plate = thermocycler.load_labware(
    "opentrons_96_wellplate_200ul_pcr_full_skirt"
)

thermocycler.open_lid()
# Pipette into plate.
thermocycler.close_lid()
thermocycler.set_lid_temperature(temperature=105)
thermocycler.set_block_temperature(
    temperature=95,
    hold_time_seconds=180,
    block_max_volume=25,
)
thermocycler.execute_profile(
    steps=[
        {"temperature": 95, "hold_time_seconds": 15},
        {"temperature": 60, "hold_time_seconds": 30},
        {"temperature": 72, "hold_time_seconds": 30},
    ],
    repetitions=35,
    block_max_volume=25,
)
thermocycler.set_block_temperature(
    temperature=4,
    block_max_volume=25,
)
thermocycler.deactivate_lid()

API 2.28 adds an optional ramp_rate to block-temperature commands.

Checks:

Magnetic Block, Flex

The Magnetic Block is passive. It has no engage or disengage methods. Separation happens by moving compatible labware onto and off the block:

magnetic_block = protocol.load_module("magneticBlockV1", "D1")

protocol.move_labware(
    labware=plate,
    new_location=magnetic_block,
    use_gripper=True,
)
protocol.delay(minutes=5)
protocol.move_labware(
    labware=plate,
    new_location="C1",
    use_gripper=True,
)

Confirm gripper compatibility and plate orientation. Avoid aspirating beads by validating settle time, aspiration side, bottom clearance, and flow rate.

Magnetic Module, OT-2

The powered Magnetic Module is OT-2-only:

magnetic_module = protocol.load_module(
    "magnetic module gen2",
    "1",
)
plate = magnetic_module.load_labware(
    "nest_96_wellplate_2ml_deep"
)

magnetic_module.engage(height_from_base=6.5)
protocol.delay(minutes=5)
magnetic_module.disengage()

Engage height is labware- and assay-specific. Do not copy an arbitrary height from another plate. The Magnetic Module is discontinued but remains supported for existing OT-2 hardware.

Absorbance Plate Reader, Flex API 2.21+

The reader is Flex-only and loads in A3–D3. Its caddy also uses the corresponding column-4 location for lid travel.

Required workflow:

  1. Load the module.
  2. Close the lid with no plate inside.
  3. Initialize the reader.
  4. Open the lid.
  5. Move a compatible plate onto the module with the Gripper.
  6. Close the lid.
  7. Read.
reader = protocol.load_module(
    module_name="absorbanceReaderV1",
    location="D3",
)
plate = protocol.load_labware(
    "corning_96_wellplate_360ul_flat",
    "C2",
)

reader.close_lid()
reader.initialize(
    mode="multi",
    wavelengths=[450, 562, 600],
)
reader.open_lid()
protocol.move_labware(
    labware=plate,
    new_location=reader,
    use_gripper=True,
)
reader.close_lid()
data = reader.read(export_filename="plate_data")

Default hardware wavelengths are 450, 562, 600, and 650 nm.

read() returns:

dict[int, dict[str, float]]

The first key is wavelength; the second is well name. In simulation every measurement is zero, and no output file is written. Guard calculations that would divide by a reading:

if not protocol.is_simulating():
    normalized = data[450]["A1"] / data[450]["H12"]

Do not move the plate-reader lid manually.

Flex Stacker, API 2.25+

Up to four Stackers attach to the right side of Flex. Each shuttle is addressed as a column-4 location:

stacker = protocol.load_module(
    module_name="flexStackerModuleV1",
    location="A4",
)

Configure one labware type per Stacker:

stacker.set_stored_labware(
    load_name="opentrons_flex_96_tiprack_200ul",
    count=5,
    lid="opentrons_flex_tiprack_lid",
)

Retrieve and move:

tip_rack = stacker.retrieve()
protocol.move_labware(
    labware=tip_rack,
    new_location="B2",
    use_gripper=True,
)

Store:

protocol.move_labware(
    labware=plate,
    new_location=stacker,
    use_gripper=True,
)
stacker.store()

fill() and empty() pause for manual loading or unloading.

Constraints:

Moving Labware and Lids

Flex automated move:

protocol.move_labware(
    labware=plate,
    new_location="C2",
    use_gripper=True,
)

Manual move:

protocol.move_labware(
    labware=plate,
    new_location="C2",
    use_gripper=False,
)

A manual move pauses for the operator. Make sure the message and run setup make the source and destination unambiguous.

API 2.23+ supports lid stacks and move_lid(). Confirm lid and labware compatibility, orientation, stack quantity, and disposal location.

Concurrent Module Actions, API 2.27+

Concurrent methods can overlap long module actions with independent pipetting:

Pattern:

  1. Start the operation and keep the returned task.
  2. Perform only physically independent commands.
  3. Wait for the task before any step that assumes completion.

Do not create concurrency merely to shorten runtime. Check deck access, vibration, thermal dependencies, and collision risk.

Module State Checklist