Official Source Ledger
Research date: 2026-07-27. Every framework claim in this skill traces to an entry below.
Re-check each source before operational use — guidelines are revised, editions change, and regional
implementation dates differ from adoption dates.
This ledger is a version baseline. It is not legal advice, an applicability determination, or a
substitute for a controlled copy held under the laboratory's document control.
Documents read directly
These were downloaded and read as full text on the research date, so the requirements encoded in
scripts/_catalog.py and summarised in references/ich-q2r2.md and
references/ich-m10-bioanalytical.md come from the primary source rather than from secondary
summaries.
ICH Q2(R2) Validation of Analytical Procedures
- Source read: https://database.ich.org/sites/default/files/ICH_Q2%28R2%29_Guideline_2023_1130.pdf
- Verified metadata: Final Version, adopted by the ICH Assembly Regulatory Members under Step 4 on
1 November 2023. Step 2 endorsement 24 March 2022. Supersedes Q2(R1) (November 2005).
- Verified detail: an error correction dated 30 November 2023 covers Table 5 (dissolution with
HPLC, reportable range linearity formulae, page 25) and Tables 6–11 (pages 26–32).
- Content taken: section structure; Table 1 (tests by measured attribute); Table 2 (reportable range
examples); recommended data for specificity, response, lower range limits, accuracy, precision,
and robustness; sections 2.1–2.5; Annex 1 and Annex 2 table inventory; the relative response factor
0.8–1.2 rule from Annex 2 Table 3.
- Licence: ICH permits use, reproduction, adaptation and distribution under a public licence provided
ICH's copyright is acknowledged. Acknowledged here and in
scripts/_catalog.py.
- Limitation: adoption is not implementation. Confirm the date from which your regional regulator
expects Q2(R2) with that regulator.
ICH M10 Bioanalytical Method Validation and Study Sample Analysis
- Source read: https://database.ich.org/sites/default/files/M10_Guideline_Step4_2022_0524.pdf
- Verified metadata: Step 4, dated 24 May 2022.
- Content taken: chromatographic criteria (section 3) — calibration levels and tolerances, QC
placement at four levels with the low/medium/high definitions, within-run and between-run accuracy
and precision design and criteria, routine-run QC pass rules, carry-over, selectivity source count,
dilution integrity, stability; ligand binding assay criteria (section 4) — calibration tolerances
including anchor point exclusion, five QC levels, run and replicate structure, accuracy and
precision criteria at LLOQ and ULOQ, and the total error criterion; incurred sample reanalysis
(section 5) including the percent-difference basis and the pass fractions.
- Verified distinction: the total error criterion (≤30%, ≤40% at LLOQ and ULOQ) appears for ligand
binding assays. No equivalent criterion was found for chromatographic assays.
- Licence: as for Q2(R2).
- Limitation: regional implementation dates differ. Confirm with the regional regulator.
ICH Q14 Analytical Procedure Development
- Source read: https://database.ich.org/sites/default/files/ICH_Q14_Guideline_2023_1116.pdf
- Content taken: section structure; the minimal versus enhanced approaches (section 2.1); the
analytical target profile (section 3) and that its formal documentation and submission is
optional; robustness and parameter ranges (section 5); established conditions (section 6.1);
lifecycle management and post-approval change (section 7); multivariate procedures (section 8).
- Adopted alongside Q2(R2) by the ICH Assembly in the same session.
- Licence: as for Q2(R2).
Documents identified but not read (paywalled)
Designation, title, and scope only. No requirement, threshold, or study design from any of these is
reproduced anywhere in this skill. Where a numeric criterion is needed, read it from an authorised
copy.
USP–NF general chapters
- Official pages:
<1220> https://doi.usp.org/USPNF/USPNF_M10975_02_01.html;
<1225> https://doi.usp.org/USPNF/USPNF_M99945_40101_01.html;
<1226> https://doi.usp.org/USPNF/USPNF_M870_03_01.html
- Verified metadata for
<1220>: incorporated into USP–NF 2022 Issue 1 on 1 November 2021,
official 1 May 2022. It brings the concepts of <1224>, <1225> and <1226> into a single
three-stage lifecycle. <1225> covers validation, particularly Stage 2 activities under <1220>;
<1226> covers verification of compendial procedures.
- Provenance limitation: this metadata came from secondary sources (publisher notices and trade
press) rather than from the USP–NF text, which is behind subscription. Marked
[confirm in USP–NF]. Confirm the current official text, revision, and any subsequent change.
- Chapters referenced by designation only, not read:
<1224>, <1010>, <621>, <711>, <1092>.
CLSI EP series
- Publisher: https://clsi.org/standards/products/method-evaluation/
- Designations and subjects recorded in
references/compendial-and-clsi.md: EP05, EP06, EP07, EP09,
EP15, EP17, EP25, EP28 (formerly C28), plus the EP17IG and EP28IG implementation guides.
- Provenance limitation: designations, titles and edition numbers were taken from clsi.org product
listings and secondary sources on the research date, not read from the documents. Every edition
number carries [confirm edition] in the reference file. Editions change; verify on clsi.org
before designing a study.
ISO standards
- ISO/IEC 17025:2017 — https://www.iso.org/standard/66912.html. Edition 3; supersedes the 2005
edition. Relevant clauses: 7.2 (selection, verification and validation of methods), 7.6
(measurement uncertainty). Not read; identified by catalogue metadata.
- ISO 15189, ISO 21748, ISO 5725 series — referenced by designation and scope only.
- Provenance limitation: ISO catalogue pages have historically refused automated access. Confirm
edition and status on iso.org or with a national member body. [confirm on iso.org]
- See this repository's
iso-standards-readiness skill and its own source ledger for the
accreditation-level treatment of these standards.
Statistical methods
The statistical procedures in references/statistics.md and scripts/_common.py are standard
published methods, not requirements of any framework:
- Incomplete beta and gamma function implementations follow the standard continued-fraction and series
algorithms; the t, chi-square and F distributions are derived from them.
- Lack-of-fit F test against pure error: standard regression ANOVA.
- Wald–Wolfowitz runs test: standard non-parametric test of randomness in a sequence of signs.
- One-way random-effects variance components with the standard unbalanced expected-mean-square
coefficient; Satterthwaite approximation for effective degrees of freedom of the total.
- Deming regression with jackknife standard errors; Passing–Bablok with the rank-based slope interval.
- Bland–Altman bias and limits of agreement.
- Two one-sided tests (TOST) for equivalence.
Implementations are verified against published quantiles and hand-checkable cases in
tests/analytical-method-validation/test_scripts.py. Where a framework prescribes a specific
statistical treatment, the framework governs — these are the general-purpose tools.
What is deliberately absent
- No numeric acceptance criteria are supplied for ICH Q2(R2) work. The guideline does not set them and
neither does this skill; they come from the specification, the analytical target profile, or
development data.
- No text, table, threshold, or study design from any USP, CLSI, or ISO document.
- No claim that a procedure is validated, a run acceptable, or an investigation closed.